CALIBRATION PROCEDURES AND INSTRUMENTAL ACCURACY ESTIMATES OF TAO TEMPERATURE, RELATIVE HUMIDITY AND
RADIATION MEASUREMENTS
H. P. Freitag, Y. Feng, L. J. Mangum, M. J. McPhaden, J. Neander, L. D. Stratton
Pacific Marine Environmental Laboratory
National Oceanic and Atmospheric Administration
7600 Sand Point Way NE
Seattle, WA 98115-0070
December 1994
Contribution No. 1589 from NOAA/Pacific Marine Environmental Laboratory
NOTICE
Mention of a commercial company or product does not constitute an endorsement by NOAA/ERL. Use of
information from this publication concerning proprietary products or the tests of such products for
publicity or advertising purposes is not authorized.
CONTENTS
-
INTRODUCTION
-
INDIVIDUAL SENSOR AND I/O BOARD CALIBRATIONS
-
Air Temperature Sensor
-
Air Temperature I/O Board
-
Sea Surface Temperature Sensor
-
Sea Surface Temperature I/O Board
-
Relative Humidity Sensors
-
Relative Humidity I/O Board
-
Shortwave Radiation Sensors
-
Shortwave Radiation I/O Board
-
Subsurface Temperature Sensors
-
SENSOR AND I/O BOARD DRIFT
-
Air Temperature Sensor
-
Air Temperature I/O Board
-
Sea Surface Temperature Sensor
-
Sea Surface Temperature I/O Board
-
Relative Humidity Sensors
-
Relative Humidity I/O Board
-
Shortwave Radiation Sensors
-
Shortwave Radiation I/O Board
-
Subsurface Temperature Sensors
-
SUMMARY
-
CONCLUSIONS
-
ACKNOWLEDGMENTS
-
REFERENCES
TABLES
-
Manufacturer, model and specifications for temperature, humidity and shortwave radiation
sensors used on TAO moorings
-
Statistics for sensor calibrations
-
Statistics for board calibrations
-
Differences between pre- and post-deployment sensor calibrations
-
Differences between pre- and post-deployment I/O board calibrations
-
Combined instrumental error for each measured parameter
FIGURES
-
Map of the tropical Pacific Ocean with location of ATLAS and PROTEUS moorings within the TAO
array shown as of December 1994
-
Schematic drawing of typical ATLAS mooring
-
Schematic drawing of typical PROTEUS mooring
-
Flow diagram of ATLAS/AMP operation
-
Absolute value of ATLAS SST sensor calibration differences at 25°C vs. the number of days
between calibrations
-
PROTEUS and ATLAS relative humidity sensor calibration differences at 90% RH vs. the number of
days the sensors were deployed at sea
Calibration Procedures and Instrumental Accuracy Estimates of TAO Temperature, Relative Humidity and
Radiation Measurements
H. Paul Freitag, Yue Feng, Linda J. Mangum, Michael J. McPhaden, LT Julia Neander, and Linda D.
Stratton
Abstract. Calibration procedures for instruments measuring air and water temperature, humidity
and shortwave radiation on Tropical Atmosphere Ocean (TAO) Array buoys are described. Initial sensor
accuracy as well as drift are quantified. Improvements in calibration procedures, where necessary, are
discussed.
INTRODUCTION
The Tropical Atmosphere Ocean (TAO) Array of moored buoys spans the tropical Pacific from longitudes
137°E to 95°W between latitudes of approximately 8°S and 8°N (Fig. 1). Moorings within
the array measure surface meteorological and upper-ocean parameters and transmit most data in real time
to shore via Service Argos. The array is part of the in-situ measurement portion of the Tropical
Ocean-Global Atmosphere (TOGA) Program, a 10-year (1985 - 1994) study of climate variability on seasonal
to interannual time scales, the most pronounced mode of which is the El Niño/Southern Oscillation
(ENSO) phenomenon (McPhaden, 1993). The TAO array is presently supported by the United States, France,
Japan, Korea and Taiwan.
Fig. 1. Map of the tropical Pacific Ocean with location of ATLAS and PROTEUS moorings within the
TAO array shown as of December 1994.
TAO moorings are designed, tested, calibrated and constructed at NOAA's Pacific Marine Environmental
Laboratory (PMEL). TAO began in 1985 as regional-scale meridional arrays spanning the equator along
110W and 165E and has steadily expanded to its present size of approximately 70 moorings. Moorings
are typically separated by 2 - 3 degrees of latitude and by 10 - 15 degrees of longitude.
The majority of TAO moorings are ATLAS moorings (Hayes et al., 1991) which measure surface wind,
air-temperature (AT), relative humidity (RH), sea-surface temperature (SST), subsurface temperature
(SBT) and pressure (P) (Fig. 2a). ATLAS moorings are designed for a nominal 1-year deployment.
Fig. 2a. Schematic drawing of typical ATLAS mooring.
At a few sites PROTEUS moorings (McPhaden et al., 1990) (Fig. 2b) are deployed, which measure and
transmit the same surface parameters as well as current profiles from Acoustic Doppler Current
Profilers (ADCPs). Recently shortwave radiation (SWR) has been added to the real-time PROTEUS
meteorological measurement suite. Surface meteorological measurements on PROTEUS moorings are made
by an AMP (Argos Meteorological Package) which is similar in design to the ATLAS. In addition,
internally recording temperature sensors (MTRs), temperature-conductivity sensors (Seacats;
western-Pacific sites only), and mechanical current meters (MCMs) measure temperature at up to 17
depths, conductivity at up to 10 depths, and current velocity at up to 7 depths, depending on the
particular mooring site. Temperature and current data from the above internally recording
instruments are not available in real time. PROTEUS moorings are designed with a nominal 6-month
deployment.
Fig. 2b. Schematic drawing of typical PROTEUS mooring.
This report covers the calibration techniques and estimated accuracies of AT, RH, SST, SWR and SBT
measurements as made on presently deployed moorings. (Other measured parameters, e.g., wind speed,
will be addressed in future reports.) The sensors used to make these measurements were purchased
from commercial vendors. The manufacturer, model number, and manufacturer's specifications for the
sensors used are listed in Table 1. The electronics hardware and software packages which digitize
and record the sensor output and pass it to the Argos transmitter were designed by PMEL's
Engineering Development Division (EDD) and constructed by TAO Project technicians.
Table 1. Manufacturer, model, and specifications for temperature, humidity, and shortwave
radiation sensors used on TAO moorings.
Measurements were typically a two stage process (Fig. 3). The environment was sampled by the sensor
and output as an analog signal (voltage, V, or resistance, R). The sensor output was converted and
stored in digital memory by input/output (I/O) boards employing either analog to digital (A/D) or
voltage to frequency (V/F) converters. Most calibrations were performed separately upon the sensor
and electronics portions of the measuring systems. An exception was the SBT cable for which sensors
and electronics were constructed as one and calibrated as a whole. Sensors and electronics were
calibrated at PMEL with the exception of SWR sensors which were calibrated by the manufacturer.
Fig. 3. Flow diagram of ATLAS/AMP operation. Boxes on left contain calibration equations relating
environmental parameter (T = temperature; RH = relative humidity; SWR = shortwave radiation) to
engineering units (V = voltage; R = resistance) output by sensors. Boxes in center contain
calibration equations relating engineering units to number (N) stored in memory by electronic I/O
boards. Calibration coefficients are denoted by a, b and c.
The ATLAS and PROTEUS mooring projects were originally headed by separate principle investigators
with separate support teams of technicians and programmers. Although the projects have since merged,
calibrations are still performed independently and calibration data bases remain separate. (For
example, ATLAS I/O board calibration coefficients are routinely scaled to be of order zero (for
bias) and 1 (for gain), but have not been normalized in this report to simplify comparison with
PROTEUS I/O boards.) Because of this separation and because of the different mooring design
lifetimes, this report will make a distinction between like sensors (AT, SST, RH) which are used on
both types of mooring. More unified and standardized procedures for calibrating ATLAS and PROTEUS
moorings are presently being instituted, however, to insure a uniform quality to all TAO
measurements.
This report has been organized into two major sections with each section subdivided by sensor or I/O
board type. The first section defines calibration equations and describes calibration methods for
each sensor or I/O board type. For each calibration the maximum residual (the largest difference
between the known input and the sensor or I/O board output as calculated using the calibration
coefficients) is computed. Focus is on the root-mean-square (RMS) of maximum residuals from the
calibration equation computed over all calibrations of a sensor or I/O board type, which may be
interpreted as an estimate of the initial sensor or board accuracy as they were deployed. The second
section deals with sensor or board drift as indicated by the difference between multiple
calibrations of the same sensor or board as it was used and reused on multiple deployments.
INDIVIDUAL SENSOR AND I/O BOARD CALIBRATIONS
Air Temperature Sensor
Air temperature measurements were made by model MP-100 humidity-temperature probes manufactured by
Rotronic Instrument Corporation of Huntington, New York. Specifications pertaining to sensor accuracy
given by the manufacturer are listed in Table 1. Sensors were calibrated at PMEL by immersing them in a
controlled water bath and measuring their output voltage at seven temperatures over a range of 14°C
to 32°C. The resultant temperature-voltage pairs were fit to a least squares linear equation. A
sample calibration is included in Appendix A. Statistics from the calibration of 31 PROTEUS and 154 ATLAS
AT sensors were quite similar (Table 2). The RMS maximum residual for both groups of sensors was an order
of magnitude smaller than the
manufacturer's stated accuracy of 0.5°C and linearity of 0.2°C. The sensor gain (coefficient
b) was essentially equal for both PROTEUS and ATLAS groups and was within 1.4% of the manufacturer's
nominal gain of 100. Sensor to sensor differences were small with the standard deviation of the gain
being only 1% of the mean. Sensor bias (coefficient a) of 0.2°C (PROTEUS) to 0.3°C (ATLAS)
is near the manufacturer's nominal bias of 0. The standard deviation as a percentage of the mean for
coefficient a is large (>100%) for both ATLAS and PROTEUS moorings because the coefficient itself
is so small.
Table 2. Statistics for sensor calibrations. M is the number of sensors calibrated. Calibration
equations are shown in Figure 3. Coefficient percent deviation is the standard deviation divided by
the mean expressed as a percent.
Table 3. Statistics for board calibrations. M is the number of boards calibrated. Calibration
equations are shown in Figure 3. Coefficient percent deviation is the standard deviation divided by
the mean expressed as a percent. Residuals have been scaled to have same units as sensors.
Air Temperature I/O Board
Air temperature I/O boards have 10-bit A/D converters and were calibrated at PMEL by applying a known
voltage and recording the digital output. These data were then fit to a least squares linear equation. A
sample calibration is included in Appendix A. Statistics from the calibration of 34 PROTEUS and 186 ATLAS
AT boards are shown in Table 3. The RMS maximum residual for the PROTEUS boards (expressed in temperature
units) was 0.019°C or half the single-bit resolution of the boards and thus as small as possible for
the given resolution. This value was half the RMS maximum residual for the AT sensor. The RMS maximum
residual for the ATLAS boards was significantly higher at 0.136°C. The difference was due to the
range of voltages over which the calibration was performed. Originally both groups of boards were
calibrated down to 0°C (0 volts) and both groups had comparable maximum residuals. It was noticed
that the maximum residual occurred almost entirely at the 0 v value. Since this was a value that would
never be experienced in the tropics the calibrations were recomputed for the PROTEUS boards after
omitting this calibration point and the maximum residuals decreased to the level shown in Table 3.
Air temperature data from both PROTEUS and ATLAS moorings archived before 1994 were computed using
calibration coefficients which used the 0 v value, but were surely more accurate than what the
0.136°C residual implies since at tropical air temperatures (18 - 32°C) calibration
residuals were typically much lower. Data archived in 1994 and after will use calibration
coefficients based on typical tropical ocean values. AT board gain (coefficient b in Table 3) was
essentially equal for both PROTEUS and ATLAS groups, despite the fact that the PROTEUS calibrations
omitted the 0 v data. This implies that the true accuracy of the ATLAS board was comparable to the
PROTEUS.
Sea Surface Temperature Sensor
Sea surface temperature measurements were made by model 46006 thermistors manufactured by Yellow Springs
Instrument Co., Inc. (YSI) of Yellow Springs, Ohio. Specifications pertaining to sensor accuracy given by
the manufacturer are listed in Table 1. Sensors were calibrated at PMEL by immersing them in a controlled
water bath and measuring their output resistance at seven temperatures over a range of 14°C to
32°C. The resultant temperature-resistance pairs were fit to a non-linear equation (Fig. 3). A sample
calibration is included in Appendix A. Statistics from the calibration of 38 PROTEUS and 269 ATLAS SST
sensors were quite similar (Table 2). The RMS maximum residual for both groups of sensors was nearly two
orders of magnitude smaller than the manufacturer's stated interchangeability of 0.2°C. Differences
between the PROTEUS and ATLAS sensors appear to be insignificant.
Sea Surface Temperature I/O Board
Sea surface temperature I/O boards converted thermistor resistance to voltage, which in turn was
converted to a frequency proportional to the voltage. This frequency was then counted for 4 seconds and
recorded. The boards were calibrated at PMEL by placing precision Vishay resistors on the boards and
recording the output counts. This procedure was repeated at seven levels which corresponded to a
temperature range of 14°C to 32°C. These data were then fit to a least squares linear equation. A
sample calibration is included in Appendix A. Statistics from the calibration of 34 PROTEUS and 196 ATLAS
SST boards are shown in Table 3. The RMS maximum residuals (expressed in temperature units) were
0.002°C (PROTEUS) and 0.004°C (ATLAS) and were of the same magnitude as the SST sensor residuals.
SST board gain (coefficient b in Table 3) was essentially equal for both PROTEUS and ATLAS groups.
Relative Humidity Sensors
Relative humidity was measured by the same sensor (Rotronic model MP-100) used to measure AT.
Specifications pertaining to sensor accuracy given by the manufacturer are listed in Table 1. Sensors
were calibrated at PMEL in a manner recommended by the manufacturer. The sensors were attached to a
calibration chamber into which a humidity standard was introduced. Both the calibration chamber and
humidity standards were supplied by the manufacturer. The humidity standards were precisely titrated
saturated-salt solutions with stated accuracies of better than 0.5% RH. Sensors were calibrated between
20% RH and 95% RH (presently modified to 50% RH to 95% RH) at 15% RH increments. A sample calibration is
included in Appendix A. Calibration chambers used on the PROTEUS and ATLAS sensors differed in that the
PROTEUS sensors were calibrated with the filter cap removed while ATLAS sensors are calibrated with the
filter cap installed. Rotronic cautions that while a clogged filter may not produce an erroneous reading,
it may significantly increase the sensor response time.
Statistics from the calibration of 33 PROTEUS and 40 ATLAS RH sensors differed more than the AT
calibrations from the same sensors (Table 2). The RMS maximum residual for PROTEUS sensors was 0.07%
RH, or about half the nominal accuracy of 2% RH, but the residual for ATLAS sensors was 2.6% RH
which exceeded the nominal accuracy by about 30%. ATLAS sensor gain (coefficient b) was 6% higher
than the nominal value of 100 while the PROTEUS gain was 3% higher than the nominal value. One
explanation for this difference would be that the ATLAS sensors did not come to equilibrium (due to
increased response time caused by clogged filters) before the calibration values were recorded. In
any case, both of the calibration residuals were large in a relative sense when compared to those
for AT and SST which were smaller than manufacturers' specifications by at least an order of
magnitude.
Relative Humidity I/O Board
Relative humidity I/O boards have 10-bit A/D converters and were calibrated at PMEL by applying a known
voltage and recording the digital output. These data were then fit to a least squares linear equation. A
sample calibration is included in Appendix A. Statistics from the calibration of 34 PROTEUS and 190 ATLAS
RH boards are shown in Table 3. The RMS maximum residuals (expressed in humidity units) were 0.17% RH
(PROTEUS) and 0.26% RH (ATLAS). Both were less than the single-bit resolution of the boards. As with the
AT boards it was found that maximum residuals tended to be found at the 0% RH (0 volts) calibration
point. This point was omitted from the PROTEUS calibration data which may account for it being smaller
than the ATLAS value. For both ATLAS and PROTEUS the RH I/O board residuals were about an order of
magnitude smaller than the sensor residuals. RH board gain (coefficient b in Table 3) was about equal for
both PROTEUS and ATLAS groups.
Shortwave Radiation Sensors
Shortwave radiation measurements (PROTEUS moorings only) were made with model PSP precision pyranometers
manufactured by the Eppley Laboratory, Inc. of Newport, Rhode Island. Specifications pertaining to sensor
accuracy given by the manufacturer are listed in Table 1. The sensors were calibrated by the manufacturer
when new and were returned for recalibration after mooring recovery. The calibration procedure entailed
comparing the sensor output at 700 W m to a
standard sensor. Quoting from Eppley's calibration report, these calibrations are "traceable to standard
self-calibrating cavity pyrheliometers in terms of the Systems Internationale des Unites (SI units),
which participated in the Seventh International Pyrheliometric Comparisons (IPCVII) at Davos, Switzerland
in October 1990." Eppley specifies that the sensors are linear to 0.5% at radiation intensities up to
1400 W m . The mean and standard deviation of
calibration coefficients from 23 calibrations performed by Eppley are given in Table 2.
Shortwave Radiation I/O Board
Shortwave radiation I/O boards have 10-bit A/D converters and were calibrated at PMEL by applying a known
voltage and recording the digital output. These data were then fit to a least squares linear equation. A
sample calibration is included in Appendix A. Statistics from the calibration of 23 PROTEUS SWR boards
are shown in Table 3. The RMS maximum residuals (expressed in radiation units) was 1.8 W m , which is only slightly larger than the resolution of the
board. By comparison, the SWR sensor's manufacturer specification of 0.5% linearity and 1% temperature
dependance imply uncertainties of 5 W m and 10 W
m, respectively, for insolation values of 1000 W
m (typical of cloudless, mid-day values in the
tropics.)
Subsurface Temperature Sensors
Subsurface temperature measurements were made by the same YSI 46006 thermistor that was used for SST.
Calibration procedures differed from those for SST in that SBT was calibrated as a total system, i.e.,
the thermistor and V/F counter were calibrated as one at 10 temperatures over a range of 6°C to
32°C. The calibration data were fit to the same non-linear equation as the SST sensor after applying
a nominal gain factor to the V/F output. A sample calibration is included in Appendix A. The RMS maximum
residual from the calibration of 3666 ATLAS SBT sensors of 0.002 C was smaller than the RMS residual for
SST sensors (Table 2). One reason for the lower SBT residual may be that the number of SBT sensors
calibrated was an order of magnitude larger than that for SST sensors and therefore less weight was given
to outliers.
SENSOR AND I/O BOARD DRIFT
When instrumentation was recovered in working condition it was returned to PMEL for post-deployment
calibration before being reused on a future deployment. Damage to some instruments by electronic
component failure, vandalism, harsh environmental conditions, loss of mooring or seal failures prevented
post-deployment calibrations in some cases. When post-deployment calibrations were made, the resultant
coefficients were compared to the pre-deployment coefficients in the following manner. A set of output
values were computed by application of the calibration equation using pre-deployment coefficients to a
set of input values. Input values were chosen so that the output values would range over normal
environmental conditions. A second set of output values were generated by application of the calibration
equation using post-deployment coefficients to the same set of input values. The first output values were
then subtracted from the second output values. Mean and RMS differences over the full output range and
for all calibration pairs are given in Table 4. Similar statistics for I/O boards are given in Table 5.
Plots of individual sensor or I/O board calibration differences are in Appendix B.
Table 4. Differences between pre- and post-deployment sensor calibration.
Table 5. Differences between pre- and post-deployment I/O bopard calibrations.
Air Temperature Sensor
RMS AT sensor calibration differences (Table 4) were similar for both PROTEUS and ATLAS groups with
values of 0.154°C (PROTEUS) and 0.168°C (ATLAS). These values were roughly four times larger than
the RMS maximum residual for AT sensors (Table 2), indicating that the differences were significant. Mean
differences were roughly 4 times smaller than RMS differences and of different sign for the two groups,
indicating that there was no preferred direction for sensor drift.
Air Temperature I/O Board
PROTEUS AT I/O boards had a RMS difference between calibration pairs of 0.046°C (Table 5), about 2.4
times the RMS maximum residual of single calibrations (Table 3), indicating that the boards drifted
measurably over time. The mean drift of 0.021°C was smaller than the single bit resolution, thus it
cannot be said that the boards had a preferred drift direction. Statistics for ATLAS I/O boards were
larger than for PROTEUS boards presumably because of the inclusion of the 0 v calibration point in their
calibration procedure (see discussion in the AT board individual calibration section above). In fact, the
RMS difference between calibration pairs, 0.113°C, was smaller than the RMS maximum residual of
single calibrations, 0.136°C, indicating no significant change between calibrations. However, PROTEUS
calibrations (which do not include the 0 v calibration point) should be regarded as a more accurate
indication of board performance.
Sea Surface Temperature Sensor
RMS SST sensor calibration differences (Table 4) were significantly larger than the individual
calibration residuals, indicating that measurable drift occurred between calibrations. RMS PROTEUS SST
sensor differences (0.014°C) were almost 5 times larger than the RMS maximum residual (Table 2),
while ATLAS sensor differences (0.030°C) were 10 times larger than RMS maximum residual. Although
ATLAS RMS calibration differences were twice as large as those for PROTEUS sensors, they equaled the
manufacturer's specified drift of 0.03°C per year. Mean differences were roughly 4 to 7 times smaller
than RMS differences indicating that there was little or no preferred direction for sensor drift. The
larger differences for ATLAS sensors may be in part due to the fact that they are deployed 1.7 times
longer than PROTEUS sensors, which is roughly the same as the ratio between ATLAS and PROTEUS RMS
difference. There was a significant, yet small, correlation (r = 0.27) between the absolute drift and the
time between calibration for ATLAS SST sensors (Fig. 4). Application of the regression slope to the mean
difference in deployment days would account for about one third of the 0.016°C difference. Other
sources for the larger drift for ATLAS SST sensors could be errors in the calibration coefficient data
base. The large number of sensors involved may increase the likelihood of a sensor ID number being
entered in error, or a sensor being modified without being noted in the data base.
Fig. 4. Absolute value of ATLAS SST sensor calibration differences at 25°C vs. the number of
days between calibrations. The dashed line is a least squares fit to the data. The correlation
coefficient for the fit is 0.27.
Sea Surface Temperature I/O Board
RMS SST I/O board calibration differences were relatively small and only 0.001°C larger than the RMS
maximum residual of the individual calibrations, indicating that SST board drift, if present, was small
relative to calibration uncertainty. SST board drifts were likewise small compared to SST sensor drifts.
Relative Humidity Sensor
RMS RH sensor calibration differences (Table 4) for both PROTEUS and ATLAS sensors were twice as large as
their respective individual calibration residuals, indicating that measurable differences occurred
between calibrations. ATLAS sensor RMS differences (4.04% RH) were about
twice as large as those for PROTEUS sensors (1.77% RH). Mean differences were positive (0.94% RH)
for ATLAS sensors, but negative ( 0.32% RH) for PROTEUS sensors. The cause of these differences is
not readily obvious. Positive differences would result if post-deployment calibrations were not
allowed to equilibrate before readings were taken. PROTEUS RH sensors were calibrated with the
filters removed, while ATLAS RH sensors were calibrated in the filters installed. As noted above,
clogged filters could significantly increase the sensor response time. Since ATLAS RH sensors were
on average deployed for relatively long periods (335 days as opposed to 181 days for PROTEUS
sensors) it could be that they became more fouled (e.g., from sea spray induced salt incrustation)
due to longer deployments. While large calibration differences did occur on ATLAS sensors which had
been deployed longer than 300 days, these large differences were both negative and positive (Fig.
5).
Fig. 5. PROTEUS and ATLAS relative humidity sensor calibration differences at 90% RH vs. the
number of days the sensors were deployed at sea.
Relative Humidity I/O Board
RMS RH I/O board calibration difference of 0.07% RH for PROTEUS boards was less than half that of the RMS
maximum residual for individual board calibrations, which indicates that no measurable drift in RH board
calibration occurred between calibrations. On the other hand, ATLAS RH board RMS difference between
calibration pairs was much larger (0.48% RH) and nearly double its RMS maximum residual for individual
board calibrations. The reason for the difference between PROTEUS and ATLAS boards is unclear, but may be
related to the fact that calibration coefficients for ATLAS boards include the voltage = 0 calibration
point, while PROTEUS boards do not. Nevertheless, RH I/O board drift for both PROTEUS and ATLAS boards
were an order of magnitude smaller than the drift of the RH sensors.
Shortwave Radiation Sensors
As noted above, shortwave radiation sensor calibrations were performed by the manufacturer. Eight sensors
have been calibrated more than once. RMS SWR sensor calibration difference was 13.7 W/m 2 (Table 4) at
700 W m, implying a relative accuracy of about 2%.
With the exception of one sensor all drifts were in the same direction, resulting in a mean difference
(11.8 W m) comparable to the RMS. The sense of the
drift is that pre-deployment calibration coefficients would underestimate radiation towards the end of
the record. Eppley does not specify drift characteristics for this sensor, but informally the
manufacturer suggests that in a tropical marine environment the drift could be as much as 2.5% per year
(George Kirk, personal communication). Drift is the result of the black lacquer coating on the sensor
fading. Our mean drift of 1.7% (11.8 W m at 700 W
m) normalized over 157 mean deployment days would
exceed 2.5% per year, but this could be due to the fact that the drift rate is maximum when the sensor is
new and should decrease as the sensor ages (ibid.).
Shortwave Radiation I/O Board
RMS SWR I/O board calibration difference was 7.7 W m , four times the individual calibration RMS maximum residual,
indicating that measurable differences occurred between calibrations. It should be noted though that the
SWR board measures relatively small voltages (order 10 æv resolution, 10 mv full scale). The RMS
calibration difference of 7.7 W m is roughly 6
bits or 60 æv which may be beyond the accuracy of the PMEL test equipment as presently used. At
this level of accuracy it would be necessary to calibrate the PMEL voltage sources against a more
accurate standard on a periodic basis and to monitor their output during each board calibration.
Subsurface Temperature Sensors
Subsurface temperature sensor calibration differences (Table 4) are significantly larger than individual
calibration uncertainties, indicating that measurable drift has occurred between calibrations. RMS
difference of 0.094°C was 3 times that for the ATLAS SST sensor and the manufacturer's specification
for annual drift. Some of the difference could be due to outliers (Appendix B) in the calibration
database. Additionally, since the SBT sensors and I/O boards are calibrated as one, some of the increased
drift could be due to the I/O boards. While the SBT V/F converters are the same ones used in the SST I/O
boards (which exhibited RMS differences of 0.005°C) they are used in a different fashion. These
include being physically mounted in a different manner, being exposed to pressures of up to 500 dbar and
using different logic to convert their output. The SBT circuitry included a precision Vishay resistor
which is used periodically to check circuit stability. Preliminary evaluation of the Vishay records
indicate that measured resistances begin to drift a few weeks after deployment, rising typically to (a
temperature equivalent of) 0.1°C, but on occasion to as large as 0.3°C. These drifts decrease on
recovery, but can be as large as 0.1°C at the time of post-deployment calibration. This apparent
drift in the SBT V/F boards has little affect on first-deployment data quality as the drift in Vishay
resistance is used to correct the output SBT values in the TAO database. Vishay resistances were not used
during calibration and therefore would affect calibration differences. Thus the apparent SBT drift of
0.094°C in Table 4 is probably too large. Second-deployment (and later) data would be adversely
affected by the overestimation of drift as their calibration coefficients would include this drift.
SUMMARY
The following generalizations and conclusions may be drawn from the above discussion.
-
Calibration residuals for sensors were generally equal to or larger than calibration residuals
for I/O boards.
-
Pre/post-deployment calibration differences (drifts) were generally equal to or larger than
calibration residuals.
-
Sensor calibration differences (drifts) were generally larger than I/O board drift.
-
AT and SST sensors performed as well as or better than specified by the manufacturer.
-
SBT sensor-board systems did not meet the specifications of the sensor manufacture. An apparent
drift in the SBT board which caused this large calibration drift probably has less affect on
the data itself. A modification of the calibration procedure may improve the error estimate.
-
While no drift specifications are published for the SWR sensor, our experience was similar to
what the manufacturer has informally suggested under conditions on TAO moorings.
-
The RH sensor drift was 4 times larger than the manufacturer's specifications. We believe the
larger than expected error was due to a combination of calibration method and environmental
fouling of the sensors.
The combined effect of board and sensor drifts shown in Table 6 is computed as (board drift2 +
sensor drift2)1/2, where it is assumed that board and sensor drifts are independent of one another.
PROTEUS AT combined error was 80% of the ATLAS value. We speculate that the ATLAS value would
approach the PROTEUS value if the ATLAS I/O board calibration coefficients were computed in a
fashion similar to the PROTEUS. PROTEUS SST and RH combined errors were half those for ATLAS.
Possible reasons for the larger ATLAS values include longer ATLAS deployments (both SST and RH),
errors in the calibration data base (SST only), and calibration procedures (RH only).
Table 6. Combined instrumental error for each measured parameter.
CONCLUSIONS
This study quantified calibration accuracy and sensor performance on ATLAS and PROTEUS moorings of the
TAO Array and in most cases found that measurement errors met the specifications of the sensor's
manufacturers. In addition, this study highlighted the need for modification in calibration procedures.
ATLAS air temperature I/O board calibrations, for example, no longer include a 0°C calibration point.
Also, the RH sensor calibration method described above is fairly time consuming and labor intensive. PMEL
technicians have found that at higher humidity values the sensor response time is much longer than that
quoted by the manufacturer. A calibration over the range 20% RH to 95% RH can take an elapsed time of 1
day and only one sensor can be used per calibration chamber. In order to decrease the time required for
sensor calibration and to hopefully decrease errors in the calibration procedure, we have modified
humidity calibrations in two ways. First, calibrations will only be performed over the range 50% RH to
95% RH since tropical humidity rarely (if ever) is below 50% RH. Secondly, a series 2500 Humidity
Generator has been purchased from Thunder Scientific Corp. of Albuquerque, New Mexico. This chamber can
accommodate up to 20 sensors at once and can be monitored and controlled by an unattended computer
program. A series of experiments will be conducted to more accurately determine sensor response time and
the effect of the filter presence and condition.
As a result of our study, ATLAS RH data based on the previous calibration techniques have been
recomputed using the manufacturer's specified coefficients as we sense that these will give better
values than the 4% error indicated in Table 4. Nevertheless, we adhere to 4% as a conservative error
for ATLAS RH measurements for the present, but expect that future calibrations will lower this error
estimate.
ACKNOWLEDGMENTS
This work was supported by NOAA's Equatorial Pacific Ocean Climate Studies (EPOCS) Program and the U.S.
TOGA Project Office. Engineering and technical support was provided by Pat McLain, Hugh Milburn, Ben
Moore, Andrew Shepherd, and David Zimmerman.
REFERENCES
Hayes, S.P., L.J. Mangum, J. Picaut, A. Sumi, and K. Takeuchi (1991): TOGA-TAO: A moored array for
real-time measurements in the tropical Pacific Ocean. Bull. Am. Meteorol. Soc., 72, 339 - 347.
McPhaden, M.J. (1993): TOGA-TAO and the 1991 93 El Niño-Southern Oscillation Event.
Oceanography, 6, 36 - 44.
McPhaden, M.J., H.B. Milburn, A.I. Nakamura, and A.J. Shepherd (1990): PROTEUS Profile Telemetry of
Upper Ocean Currents. In: Proceedings of the Marine Tech. Soc. Conference, September 25-28,
1990, The Marine Technology Society, Washington, D.C., 353 - 357.
Return to Table of Contents
|